◪ Power supply: 220V 50HZ
◪ Sterilization method: high temperature and high pressure steam
◪ Volume: 18L
◪ Rated operating temperature: 134℃
◪ Rated working pressure: 0.21Mpa
◪ Temperature and pressure control range: 105℃~134℃/0.05Mpa~0.22Mpa
◪ Sterilization time range: 0~60 minutes can be adjusted
◪ Inner container and sealed door: SUS304 stainless steel liner once casting
◪ Pressure controller: control accuracy 0.005Mpa
◪ Safety valve opening and closing range: 0.23Mpa~0.26Mpa
◪ Temperature uniformity: ≤±0.5℃
◪ Over temperature automatic protection device: over the set temperature, the system automatically terminates the work process
◪ Door safety interlock device: only the door closed in place, to start the work procedure, the chamber pressure, the door cannot be opened
◪ Super pressure automatic deflated safety valve: over the set pressure, the safety valve open release pressure
◪ Input power: 1800W
◪ Working environment temperature: 0 ~ 40℃
◪ The whole pulse vacuum drying vacuum degree: up to -0.8-0.9bar
◪ Equipped with B& D test and vacuum test program, which is used for steam penetration test
◪ Residual temperature after drying: no more than 0.2%
◪ Jet steam generator, temperature uniformity in the pot
◪ Full computer control, automatic operation, automatic fault diagnosis function
◪ Automatic data printing in disinfection process
Contact